Left ventricular lead location and the risk of ventricular arrhythmias in the MADIT-CRT trial

Eur Heart J. 2013 Jan;34(3):184-90. doi: 10.1093/eurheartj/ehs334. Epub 2012 Oct 10.

Abstract

Aims: We aimed to evaluate the influence of left ventricular (LV) lead position on the risk of ventricular tachyarrhythmias in cardiac resynchronization therapy (CRT) patients.

Methods and results: Left ventricular (LV) lead position was evaluated by biplane coronary venograms and anterior/posterior, lateral chest X-rays in patients enrolled in the Multicenter Automatic Defibrillator Implantation Trial - Cardiac Resynchronization Therapy (MADIT-CRT). The LV lead location could be defined in 797 of 1089 patients (73%). The LV lead was placed at the LV apex in 110 (14%) patients, in the anterior position in 146 (18%), in the lateral position in 448 (56%), and in the posterior position in 93 (12%) patients. After adjustment for clinical covariates, lateral or posterior lead location was associated with significantly lower risk of ventricular tachycardia (VT)/ventricular fibrillation (VF) [hazard ratio (HR) = 0.57, 95% confidence interval (CI): 0.38-0.85; P = 0.006] when compared with an anterior lead location. Patients with anterior lead position had similar risk of VT/VF as patients with implantable cardioverter defibrillator (ICD)-only (HR = 1.04, 95% CI: 0.72-1.81; P = 0.837). There was no difference in the risk of mortality between posterior or lateral and anterior LV lead locations.

Conclusion: Cardiac resynchronization therapy with posterior or lateral LV lead position is associated with decreased risk of arrhythmic events in comparison with anterior lead location and ICD-only patients. There is no evidence for increased risk of VT/VF episodes associated with CRT.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cardiac Resynchronization Therapy Devices / adverse effects*
  • Cardiomyopathies / therapy*
  • Death, Sudden, Cardiac / etiology*
  • Defibrillators, Implantable / adverse effects
  • Female
  • Heart Failure / therapy*
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Prospective Studies
  • Risk Factors
  • Tachycardia, Ventricular / etiology*
  • Ventricular Fibrillation / etiology*