Multiple-breath washout measurements can be significantly shortened in children

Thorax. 2013 Jun;68(6):586-7. doi: 10.1136/thoraxjnl-2012-202345. Epub 2012 Oct 31.

Abstract

Multiple-breath washout (MBW)-derived lung clearance index (LCI) is a sensitive measure of ventilation inhomogeneity in patients with cystic fibrosis (CF), but LCI measurement is time consuming. We systematically assessed ways to shorten LCI measurements. In 68 school-aged children (44 with mild CF lung disease) three standard nitrogen (N2) MBWs were applied. We assessed repeatability and diagnostic performance of (1) LCI measured earlier from three MBW runs and (2) LCI measured at complete MBW (1/40th of starting N2 concentration) from two runs only. Compared with the standard LCI from three complete MBW runs, the new LCI based on three N2MBW runs until 1/20th, or two complete runs until 1/40th, provided similar or better repeatability as well as sensitivity and specificity for CF lung disease. Alternative ways to measure LCI reduced test duration in children with CF by 30% and 41%, respectively. LCI measurements can be reliably shortened in children. These new MBW protocols may advance the transition of LCI from research into clinical settings.

Keywords: Cystic Fibrosis; Lung Physiology; Paediatric Lung Disaese.

Publication types

  • Letter
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Breath Tests / methods*
  • Child
  • Cystic Fibrosis / diagnosis*
  • Cystic Fibrosis / physiopathology
  • Exhalation
  • Female
  • Forced Expiratory Volume
  • Humans
  • Lung Diseases / diagnosis*
  • Lung Diseases / physiopathology
  • Male
  • Nitrogen Dioxide / analysis
  • ROC Curve

Substances

  • Nitrogen Dioxide