Images of radiation-sensitive specimens obtained by electron microscopy suffer a reduction in quality beyond that expected from radiation damage alone due to electron beam-induced charging or movement of the specimen. For biological specimens, charging and movement are most severe when they are suspended in an insulating layer of vitreous ice, which is otherwise optimal for preserving hydrated specimens in a near native state. We image biological specimens, including a single particle protein complex and a lipid-enveloped virus in thin, vitreous ice films over suspended sheets of unmodified graphene. We show that in such preparations, the charging of ice, as assessed by electron-optical perturbation of the imaging beam, is eliminated. We also use the same specimen supports to record high resolution images at liquid nitrogen temperature of monolayer paraffin crystals grown over graphene.
Keywords: Beam induced specimen movement; Charging; Cryomicroscopy; Graphene; Radiation damage.
Copyright © 2013 The Authors. Published by Elsevier Inc. All rights reserved.