Patient-specific quality assurance for the delivery of (60)Co intensity modulated radiation therapy subject to a 0.35-T lateral magnetic field

Int J Radiat Oncol Biol Phys. 2015 Jan 1;91(1):65-72. doi: 10.1016/j.ijrobp.2014.09.008. Epub 2014 Oct 25.

Abstract

Purpose: This work describes a patient-specific dosimetry quality assurance (QA) program for intensity modulated radiation therapy (IMRT) using ViewRay, the first commercial magnetic resonance imaging-guided RT device.

Methods and materials: The program consisted of: (1) a 1-dimensional multipoint ionization chamber measurement using a customized 15-cm(3) cube-shaped phantom; (2) 2-dimensional (2D) radiographic film measurement using a 30- × 30- × 20-cm(3) phantom with multiple inserted ionization chambers; (3) quasi-3D diode array (ArcCHECK) measurement with a centrally inserted ionization chamber; (4) 2D fluence verification using machine delivery log files; and (5) 3D Monte Carlo (MC) dose reconstruction with machine delivery files and phantom CT.

Results: Ionization chamber measurements agreed well with treatment planning system (TPS)-computed doses in all phantom geometries where the mean ± SD difference was 0.0% ± 1.3% (n=102; range, -3.0%-2.9%). Film measurements also showed excellent agreement with the TPS-computed 2D dose distributions where the mean passing rate using 3% relative/3 mm gamma criteria was 94.6% ± 3.4% (n=30; range, 87.4%-100%). For ArcCHECK measurements, the mean ± SD passing rate using 3% relative/3 mm gamma criteria was 98.9% ± 1.1% (n=34; range, 95.8%-100%). 2D fluence maps with a resolution of 1 × 1 mm(2) showed 100% passing rates for all plan deliveries (n=34). The MC reconstructed doses to the phantom agreed well with planned 3D doses where the mean passing rate using 3% absolute/3 mm gamma criteria was 99.0% ± 1.0% (n=18; range, 97.0%-100%), demonstrating the feasibility of evaluating the QA results in the patient geometry.

Conclusions: We developed a dosimetry program for ViewRay's patient-specific IMRT QA. The methodology will be useful for other ViewRay users. The QA results presented here can assist the RT community to establish appropriate tolerance and action limits for ViewRay's IMRT QA.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Abdominal Neoplasms / radiotherapy
  • Breast Neoplasms / radiotherapy
  • Calibration
  • Cobalt Radioisotopes / therapeutic use*
  • Female
  • Film Dosimetry
  • Humans
  • Lung Neoplasms / radiotherapy
  • Magnetic Fields*
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Male
  • Monte Carlo Method
  • Neoplasms / radiotherapy*
  • Phantoms, Imaging*
  • Quality Assurance, Health Care / standards*
  • Radiotherapy Dosage
  • Radiotherapy, Image-Guided / instrumentation
  • Radiotherapy, Image-Guided / methods*
  • Radiotherapy, Intensity-Modulated / instrumentation
  • Radiotherapy, Intensity-Modulated / methods*
  • Radiotherapy, Intensity-Modulated / standards
  • Rectal Neoplasms / radiotherapy
  • Reproducibility of Results
  • Urinary Bladder Neoplasms / radiotherapy

Substances

  • Cobalt Radioisotopes