Frameless, surface imaging guided radiosurgery (SIG-RS) is a novel platform for stereotactic radiosurgery (SRS) wherein patient positioning is monitored in real-time through infra-red camera tracking of facial topography. Here we describe our initial clinical experience with SIG-RS for the treatment of benign neoplasms of the skull base. We identified 48 patients with benign skull base tumors consecutively treated with SIG-RS at a single institution between 2009 and 2011. Patients were diagnosed with meningioma (n = 22), vestibular schwannoma (n = 20), or nonfunctional pituitary adenoma (n = 6). Local control and treatment-related toxicity were retrospectively assessed. Median follow-up was 65 months (range 61-72 months). Prescription doses were 12-13 Gy in a single fraction (n = 18), 8 Gy × 3 fractions (n = 6), and 5 Gy × 5 fractions (n = 24). Actuarial tumor control rate at 5 years was 98%. No grade ≥3 treatment-related toxicity was observed. Grade ≤2 toxicity was associated with symptomatic lesions (p = 0.049) and single fraction treatment (p = 0.005). SIG-RS for benign skull base tumors produces clinical outcomes comparable to conventional frame-based SRS techniques while enhancing patient comfort.
Keywords: Benign skull base tumor; Frameless radiosurgery; Image guided radiosurgery; SIG-RS; Stereotactic radiosurgery; Surface imaging guided radiosurgery.