Detection of plasma MMP-9 within minutes. Unveiling some of the clues to develop fast and simple electrochemical magneto-immunosensors

Biosens Bioelectron. 2018 Sep 15:115:45-52. doi: 10.1016/j.bios.2018.05.020. Epub 2018 May 11.

Abstract

Magnetic beads (MB) have been extensively used to produce sensitive and efficient electrochemical magneto-immunosensors. However, MB effective handling requires training, and MB washing after each incubation step is time consuming and contributes to raise result variability. Consequently, most of the electrochemical magneto-immunosensors reported to date, which entailed relatively long and complex multi-step procedures, would be difficult to carry out at point-of-care (POC) settings or by laypersons. For this reason, here we targeted the development of a simplified detection path, which is fast and simple enough to be operated at a POC setting, sufficiently efficient to provide analyte quantitation comparable to classical diagnostic methods, and dependent on minimal technical requirements to facilitate method global exploitation. As a proof-of-concept, we optimized an extremely simple, fast and efficient electrochemical magneto-immunosensor for detection of matrix metalloproteinase 9 (MMP-9). To accomplish this, we optimized MB immunomodification, produced an immunomodified Poly-HRP signal amplifier, developed a single-step magneto-immunoassay, and optimized electrochemical detection using a multiplexed magnetic holder and a ready-to-use commercial substrate solution. The sensor was finally calibrated by detecting MMP-9 in clinical samples. This electrochemical magneto-immunosensor detected MMP-9 in just 12-15 min, displaying linear response between 0.03 and 2 ng mL-1 of MMP-9, limits of detection (LOD) and quantification (LOQ) of 13 pg mL-1 and 70 pg mL-1, respectively, %CV< 6%, and accurate quantification of MMP-9 in patient plasma samples. These results were comparable to those afforded by a 5-h reference ELISA that used the same antibodies, confirming the applicability of our simplified method.

Keywords: Electrochemical detection; Electrochemical magneto-immunosensor; Magnetic beads (MB); Matrix metalloproteinase 9 (MMP-9); Poly-HRP signal amplifier; Stroke biomarker.

MeSH terms

  • Antibodies / blood
  • Antibodies / chemistry
  • Biosensing Techniques*
  • Electrochemical Techniques*
  • Enzyme-Linked Immunosorbent Assay
  • Humans
  • Limit of Detection
  • Magnetics
  • Matrix Metalloproteinase 9 / blood*
  • Matrix Metalloproteinase 9 / chemistry

Substances

  • Antibodies
  • Matrix Metalloproteinase 9