In silico assessment of the effects of material on stent deployment

Proc IEEE Int Symp Bioinformatics Bioeng. 2017 Oct:2017:462-467. doi: 10.1109/BIBE.2017.00-11. Epub 2018 Jan 11.

Abstract

Coronary stents are expandable scaffolds that are used to widen occluded diseased arteries and restore blood flow. Because of the strain they are exposed to and forces they must resist as well as the importance of surface interactions, material properties are dominant. Indeed, a common differentiating factors amongst commercially available stents is their material. Several performance requirements relate to stent materials including radial strength for adequate arterial support post-deployment. This study investigated the effect of the stent material in three finite element models using different stents made of: (i) Cobalt-Chromium (CoCr), (ii) Stainless Steel (SS316L), and (iii) Platinum Chromium (PtCr). Deployment was investigated in a patient specific arterial geometry, created based on a fusion of angiographic data and intravascular ultrasound images. In silico results show that: (i) the maximum von Mises stress occurs for the CoCr, however the curved areas of the stent links present higher stresses compared to the straight stent segments for all stents, (ii) more areas of high inner arterial stress exist in the case of the CoCr stent deployment, (iii) there is no significant difference in the percentage of arterial stress volume distribution among all models.

Keywords: finite element model; material; mathematical model; stent.