Performance of fast-ion loss diagnostic on EAST

Rev Sci Instrum. 2018 Oct;89(10):10I144. doi: 10.1063/1.5038782.

Abstract

The scintillator-based detector for fast-ion loss measurements has been installed on EAST. To obtain high temporal resolution for fast-ion loss diagnostics, fast photomultiplier tube systems have been developed which can supply the complementary measurements to the previous image system with good energy and pitch resolution by using a CCD camera. By applying the rotatable platform, the prompt losses of beam-ions can be measured in normal and reverse magnetic field. The thick-target bremsstrahlung occurring in the stainless steel shield with energetic electrons can produce X-rays, which will strike on the scintillator based detector. To understand this interference on fast-ion loss signals, the effects of energetic electrons on the scintillator-based detector are studied, including runaway electrons in the plasma ramping-up phase and fast electrons accelerated by the lower hybrid wave.