Estimating PTV Margins in Head and Neck Stereotactic Ablative Radiation Therapy (SABR) Through Target Site Analysis of Positioning and Intrafractional Accuracy

Int J Radiat Oncol Biol Phys. 2020 Jan 1;106(1):185-193. doi: 10.1016/j.ijrobp.2019.09.010. Epub 2019 Sep 30.

Abstract

Purpose: Recurrent or previously irradiated head and neck cancers (HNC) are therapeutically challenging and may benefit from high-dose, highly accurate radiation techniques, such as stereotactic ablative radiation therapy (SABR). Here, we compare set-up and positioning accuracy across HNC subsites to further optimize the treatment process and planning target volume (PTV) margin recommendations for head and neck SABR.

Methods and materials: We prospectively collected data on 405 treatment fractions across 79 patients treated with SABR for recurrent/previously irradiated HNC. First, interfractional error was determined by comparing ExacTrac x-ray to the treatment plan. Patients were then shifted and residual error was measured with repeat x-ray. Next, cone beam computed tomography (CBCT) was compared with ExacTrac for positioning agreement, and final shifts were applied. Lastly, intrafractional error was measured with x-ray before each arc. Results were stratified by treatment site into skull base, neck/parotid, and mucosal.

Results: Most patients (66.7%) were treated to 45 Gy in 5 fractions (range, 21-47.5 Gy in 3-5 fractions). The initial mean ± standard deviation interfractional errors were -0.2 ± 1.4 mm (anteroposterior), 0.2 ± 1.8 mm (craniocaudal), and -0.1 ± 1.7 mm (left-right). Interfractional 3-dimensional vector error was 2.48 ± 1.44, with skull base significantly lower than other sites (2.22 vs 2.77; P = .0016). All interfractional errors were corrected to within 1.3 mm and 1.8°. CBCT agreed with ExacTrac to within 3.6 mm and 3.4°. CBCT disagreements and intrafractional errors of >1 mm or >1° occurred at significantly lower rates in skull base sites (CBCT: 16.4% vs 50.0% neck, 52.0% mucosal, P < .0001; intrafractional: 22.0% vs 48.7% all others, P < .0001). Final PTVs were 1.5 mm (skull base), 2.0 mm (neck/parotid), and 1.8 mm (mucosal).

Conclusions: Head and neck SABR PTV margins should be optimized by target site. PTV margins of 1.5 to 2 mm may be sufficient in the skull base, whereas 2 to 2.5 mm may be necessary for neck and mucosal targets. When using ExacTrac, skull base sites show significantly fewer uncertainties throughout the treatment process, but neck/mucosal targets may require the addition of CBCT to account for positioning errors and internal organ motion.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Aged
  • Cone-Beam Computed Tomography / methods
  • Dose Fractionation, Radiation
  • Head and Neck Neoplasms / diagnostic imaging
  • Head and Neck Neoplasms / pathology
  • Head and Neck Neoplasms / radiotherapy*
  • Humans
  • Immobilization / methods
  • Middle Aged
  • Neoplasm Recurrence, Local / diagnostic imaging
  • Neoplasm Recurrence, Local / pathology
  • Neoplasm Recurrence, Local / radiotherapy*
  • Parotid Neoplasms / diagnostic imaging
  • Parotid Neoplasms / radiotherapy
  • Patient Positioning
  • Prospective Studies
  • Quality Improvement
  • Radiosurgery / methods*
  • Radiotherapy Planning, Computer-Assisted / methods
  • Radiotherapy Setup Errors*
  • Radiotherapy, Image-Guided / methods
  • Re-Irradiation / methods
  • Skull Base Neoplasms / diagnostic imaging
  • Skull Base Neoplasms / radiotherapy