Background: Molecular assays based on reverse transcription-loop-mediated isothermal amplification (RT-LAMP) may be useful for rapid diagnosis of the severe acute respiratory syndrome Coronavirus-2 (SARS-CoV-2) because of the easy performance and the option to bypass RNA extraction.
Objectives: This study was designed to evaluate the clinical performance of the CE-labeled variplexTM real time SARS-CoV-2 RT-LAMP assay in comparison to commercial RT-PCRs.
Study design: RNA extracted from pharyngeal swabs was tested by variplex™ RT-LAMP and Corman's LightMix™ E gene RT-PCR as reference. Samples of respiratory secretions from Coronavirus infection disease (COVID-19) and negative control patients were analyzed by variplex™ without RNA extraction and tested in parallel with the Allplex™ and VIASURE BD MAX RT-PCRs.
Results: Using isolated RNA variplex™ RT-LAMP showed a sensitivity of 75 % compared to LightMix E gene RT-PCR but contrary to the latter it produced no false-positive results. For the evaluation of samples from respiratory secretions concordance analysis showed only a moderate agreement between the variplex™ RT-LAMP conducted on unprocessed samples and Allplex™ and VIASURE RT-PCRs (Cohen's κ ranging from 0.52-0.56). Using the approach to define a sample as true-positive when at least two assays gave a positive result the clinical sensitivities were as follows: 76.3 % for variplex™, 84.2 % for Allplex™ and 68.4 % for VIASURE. However, when results of RT-PCR and RT-LAMP were combined diagnostic sensitivity was increased to 92-100 %.
Conclusion: The variplex RT-LAMP may serve as a rapid test to be combined with a RT-PCR assay to increase the diagnostic accuracy in patients with suspected COVID-19 infection.
Keywords: RT-LAMP; RT-PCR; Rapid diagnostics; SARS-CoV-2.