Is cerebral perfusion maintained during full and partial resuscitative endovascular balloon occlusion of the aorta in hemorrhagic shock conditions?

J Trauma Acute Care Surg. 2021 Jul 1;91(1):40-46. doi: 10.1097/TA.0000000000003124.

Abstract

Background: Partial resuscitative endovascular balloon occlusion of the aorta (pREBOA) is a technology that occludes aortic flow and allows for controlled deflation and restoration of varying distal perfusion. Carotid flow rates (CFRs) during partial deflation are unknown. Our aim was to measure CFR with the different pREBOA balloon volumes and correlate those to the proximal mean arterial pressure (PMAP) and a handheld pressure monitoring device (COMPASS; Mirador Biomedical, Seattle, WA).

Methods: Ten swine underwent a hemorrhagic injury model with carotid and iliac arterial pressures monitored via arterial lines. Carotid and aortic flow rates were monitored with Doppler flow probes. A COMPASS was placed to monitor proximal pressure. The pREBOA was inflated for 15 minutes then partially deflated for an aortic flow rate of 0.7 L/min for 45 minutes. It was then completely deflated. Proximal mean arterial pressures and CFR were measured, and correlation was evaluated. Correlation between CRF and COMPASS measurements was evaluated.

Results: Carotid flow rate increased 240% with full inflation. Carotid flow rate was maintained at 100% to 150% of baseline across a wide range of partial deflation. After full deflation, CFR transiently decreased to 45% to 95% of baseline. There was strong positive correlation (r > 0.85) between CFR and PMAP after full inflation, and positive correlation with partial inflation (r > 0.7). Carotid flow rate had strong correlation with the COMPASS with full REBOA (r > 0.85) and positive correlation with pREBOA (r > 0.65).

Conclusion: Carotid flow rate is increased in a hemorrhagic model during full and partial inflation of the pREBOA and correlates well with PMAP. Carotid perfusion appears maintained across a wide range of pREBOA deflation and could be readily monitored with a handheld portable COMPASS device instead of a standard arterial line setup.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Aorta / surgery
  • Balloon Occlusion / adverse effects*
  • Balloon Occlusion / instrumentation
  • Blood Flow Velocity
  • Carotid Arteries / physiology
  • Cerebrovascular Circulation*
  • Disease Models, Animal
  • Hemostatic Techniques / adverse effects*
  • Hemostatic Techniques / instrumentation
  • Humans
  • Male
  • Resuscitation / adverse effects*
  • Resuscitation / instrumentation
  • Resuscitation / methods
  • Shock, Hemorrhagic / therapy*
  • Swine