Three-micrometer-diameter needle electrode with an amplifier for extracellular in vivo recordings

Proc Natl Acad Sci U S A. 2021 Apr 20;118(16):e2008233118. doi: 10.1073/pnas.2008233118.

Abstract

Microscale needle-electrode devices offer neuronal signal recording capability in brain tissue; however, using needles of smaller geometry to minimize tissue damage causes degradation of electrical properties, including high electrical impedance and low signal-to-noise ratio (SNR) recording. We overcome these limitations using a device assembly technique that uses a single needle-topped amplifier package, called STACK, within a device of ∼1 × 1 mm2 Based on silicon (Si) growth technology, a <3-µm-tip-diameter, 400-µm-length needle electrode was fabricated on a Si block as the module. The high electrical impedance characteristics of the needle electrode were improved by stacking it on the other module of the amplifier. The STACK device exhibited a voltage gain of >0.98 (-0.175 dB), enabling recording of the local field potential and action potentials from the mouse brain in vivo with an improved SNR of 6.2. Additionally, the device allowed us to use a Bluetooth module to demonstrate wireless recording of these neuronal signals; the chronic experiment was also conducted using STACK-implanted mice.

Keywords: MOSFET; microelectrode; neural recording.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Action Potentials / physiology
  • Animals
  • Brain / physiology
  • Electric Impedance
  • Electrodes, Implanted / adverse effects
  • Electroencephalography / instrumentation*
  • Electroencephalography / methods
  • Electrophysiology / instrumentation*
  • Electrophysiology / methods*
  • Equipment Design
  • Male
  • Mice
  • Microelectrodes / adverse effects
  • Neurons / physiology
  • Signal-To-Noise Ratio