Purpose: We investigated the performance of the simplified knowledge-based plans (KBPs) in stereotactic body radiotherapy (SBRT) with volumetric-modulated arc therapy (VMAT) for lung cancer.
Materials and methods: For 50 cases who underwent SBRT, only three structures were registered into knowledge-based model: total lung, spinal cord, and planning target volume. We performed single auto-optimization on VMAT plans in two steps: 19 cases used for the model training (closed-loop validation) and 16 new cases outside of training set (open-loop validation) for TrueBeam (TB) and Halcyon (Hal) linacs. The dosimetric parameters were compared between clinical plans (CLPs) and KBPs: CLPclosed, KBPclosed-TB and KBPclosed-Hal in closed-loop validation, CLPopen, KBPopen-TB and KBPopen-Hal in open-loop validation.
Results: All organs at risk were comparable between CLPs and KBPs except for contralateral lung: V5 of KBPs was approximately 3%-7% higher than that of CLPs. V20 of total lung for KBPs showed comparable to CLPs; CLPclosed vs. KBPclosed-TB and CLPclosed vs. KBPclosed-Hal: 4.36% ± 2.87% vs. 3.54% ± 1.95% and 4.36 ± 2.87% vs. 3.54% ± 1.94% (P = 0.54 and 0.54); CLPopen vs. KBPopen-TB and CLPopen vs. KBPopen-Hal: 4.18% ± 1.57% vs. 3.55% ± 1.27% and 4.18% ± 1.57% vs. 3.67% ± 1.26% (P = 0.19 and 0.27). CI95 of KBPs with both linacs was superior to that of the CLP in closed-loop validation: CLPclosed vs. KBPclosed-TB vs. KBPclosed-Hal: 1.32% ± 0.12% vs. 1.18% ± 0.09% vs. 1.17% ± 0.06% (P < 0.01); and open-loop validation: CLPopen vs. KBPopen-TB vs. KBPopen-Hal: 1.22% ± 0.09% vs. 1.14% ± 0.04% vs. 1.16% ± 0.05% (P ≤ 0.01).
Conclusions: The simplified KBPs with limited number of structures and without planner intervention were clinically acceptable in the dosimetric parameters for lung VMAT-SBRT planning.
Keywords: Knowledge-based plan; non-small cell lung cancer; plan quality; stereotactic body radiotherapy.
Copyright: © 2021 Journal of Medical Physics.