Comparison of lesion characteristics between conventional and high-power short-duration ablation using contact force-sensing catheter in patients with paroxysmal atrial fibrillation

BMC Cardiovasc Disord. 2021 Aug 9;21(1):387. doi: 10.1186/s12872-021-02196-y.

Abstract

Background: Transmural lesion creation is essential for effective atrial fibrillation (AF) ablation. Lesion characteristics between conventional energy and high-power short-duration (HPSD) setting in contact force-guided (CF) ablation for AF remained unclear.

Methods: Eighty consecutive AF patients who received CF with conventional energy setting (power control: 25-30 W, force-time integral = 400 g s, n = 40) or with HPSD (power control: 40-50 W, 10 s, n = 40) ablation were analyzed. Of them, 15 patients in each conventional and HPSD group were matched by age and gender respectively for ablation lesions analysis. Type A and B lesions were defined as a lesion with and without significant voltage reduction after ablation, respectively. The anatomical distribution of these lesions and ablation outcomes among the 2 groups were analyzed.

Results: 1615 and 1724 ablation lesions were analyzed in the conventional and HPSD groups, respectively. HPSD group had a higher proportion of type A lesion compared to conventional group (P < 0.01). In the conventional group, most type A lesions were at the right pulmonary vein (RPV) posterior wall (50.2%) whereas in the HPSD group, most type A lesions were at the RPV anterior wall (44.0%) (P = 0.04). The procedure time and ablation time were significantly shorter in the HPSD group than that in the conventional group (91.0 ± 12.1 vs. 124 ± 14.2 min, P = 0.03; 30.7 ± 19.2 vs. 57.8 ± 21 min, P = 0.02, respectively). At a mean follow-up period of 11 ± 1.4 months, there were 13 and 7 patients with recurrence in conventional and HPSD group respectively (P = 0.03).

Conclusion: Optimal ablation lesion characteristics and distribution after conventional and HPSD ablation differed significantly. HPSD ablation had shorter ablation time and lower recurrence rate than did conventional ablation.

Keywords: Atrial fibrillation; Contact force-guided ablation; Force time integral; High-power short-duration ablation; Pulmonary vein isolation; Voltage.

Publication types

  • Comparative Study

MeSH terms

  • Age Factors
  • Anti-Arrhythmia Agents / therapeutic use
  • Atrial Fibrillation / drug therapy
  • Atrial Fibrillation / surgery*
  • Case-Control Studies
  • Catheter Ablation / instrumentation
  • Catheter Ablation / methods*
  • Catheter Ablation / statistics & numerical data
  • Electrophysiologic Techniques, Cardiac
  • Female
  • Humans
  • Male
  • Middle Aged
  • Operative Time
  • Pulmonary Veins / injuries*
  • Pulmonary Veins / physiopathology
  • Recurrence
  • Sex Factors
  • Smart Materials
  • Time Factors
  • Treatment Outcome

Substances

  • Anti-Arrhythmia Agents
  • Smart Materials