Objective: Proton range uncertainties can compromise the effectiveness of proton therapy treatments. Water equivalent path length (WEPL) assessment by flat panel detector proton radiography (FP-PR) can provide means of range uncertainty detection. Since WEPL accuracy intrinsically relies on the FP-PR calibration parameters, the purpose of this study is to establish an optimal calibration procedure that ensures high accuracy of WEPL measurements. To that end, several calibration settings were investigated.
Approach: FP-PR calibration datasets were obtained simulating PR fields with different proton energies, directed towards water-equivalent material slabs of increasing thickness. The parameters investigated were the spacing between energy layers (ΔE) and the increment in thickness of the water-equivalent material slabs (ΔX) used for calibration. 30 calibrations were simulated, as a result of combining ΔE = 9, 7, 5, 3, 1 MeV and ΔX = 10, 8, 5, 3, 2, 1 mm. FP-PRs through a CIRS electron density phantom were simulated, and WEPL images corresponding to each calibration were obtained. Ground truth WEPL values were provided by range probing multi-layer ionization chamber simulations on each insert of the phantom. Relative WEPL errors between FP-PR simulations and ground truth were calculated for each insert. Mean relative WEPL errors and standard deviations across all inserts were computed for WEPL images obtained with each calibration.
Main results: Large mean and standard deviations were found in WEPL images obtained with large ΔEvalues (ΔE = 9 or 7 MeV), for any ΔX. WEPL images obtained with ΔE ≤ 5 MeV and ΔX ≤ 5 mm resulted in a WEPL accuracy with mean values within ±0.5% and standard deviations around 1%.
Significance: An optimal FP calibration in the framework of this study was established, characterized by 3 MeV ≤ ΔE ≤ 5 MeV and 2 mm ≤ ΔX ≤ 5 mm. Within these boundaries, highly accurate WEPL acquisitions using FP-PR are feasible and practical, holding the potential to assist future online range verification quality control procedures.
Keywords: adaptive proton therapy; calibration; flat panel detector; proton radiography; range uncertainties; water equivalent path length.
Creative Commons Attribution license.