Bias-voltage dependent operational characteristics of a fully spectroscopic pixelated cadmium telluride detector system within an experimental benchtop x-ray fluorescence imaging setup

Biomed Phys Eng Express. 2021 Dec 7;8(1):10.1088/2057-1976/ac3d9c. doi: 10.1088/2057-1976/ac3d9c.

Abstract

Commercially available fully spectroscopic pixelated cadmium telluride (CdTe) detector systems have been adopted lately for benchtop x-ray fluorescence (XRF) imaging/computed tomography (XFCT) of objects containing metal nanoprobes such as gold nanoparticles (GNPs). To date, however, some important characteristics of such detector systems under typical operating conditions of benchtop XRF/XFCT imaging systems are not well known. One important but poorly studied characteristic is the effect of detector bias-voltage on photon counting efficiency, energy resolution, and the resulting material detection limit. In this work, therefore, we investigated these characteristics for a commercial pixelated detector system adopting a 1-mm-thick CdTe sensor (0.25-mm pixel-pitch), known as HEXITEC, incorporated into an experimental benchtop cone-beam XFCT system with parallel-hole detector collimation. The detector system, operated at different bias-voltages, was used to acquire the gold XRF/Compton spectra from 1.0 wt% GNP-loaded phantom irradiated with 125 kVp x-rays filtered by 1.8-mm Tin. At each bias-voltage, the gold XRF signal, and the full-width-at-half-maximum at gold Kα2XRF peak (∼67 keV) provided photon counting efficiency and energy resolution, respectively. Under the current experimental conditions, the detector photon counting efficiency and energy resolution improved with increasing bias-voltage by ∼41 and ∼29% at -300V; ∼54 and ∼35% at -500V, respectively, when compared to those at -100V. Consequently, the GNP detection limit improved by ∼26% at -300V and ∼30% at -500V. Furthermore, the homogeneity of per-pixel energy resolution within the collimated detector area improved by ∼34% at -300V and ∼54% at -500V. These results suggested the gradual improvements in the detector performance with increasing bias-voltage up to -500V. However, at and beyond -550V, there were no discernible improvements in photon counting efficiency and energy resolution. Thus, the bias-voltage range of -500 to -550V was found optimal under the current experimental conditions that are considered typical of benchtop XRF/XFCT imaging tasks.

Keywords: benchtop x-ray fluorescence imaging; detector bias-voltage optimization; gold nanoparticles; pixelated cadmium telluride detectors.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Cadmium Compounds*
  • Gold / chemistry
  • Metal Nanoparticles* / chemistry
  • Optical Imaging
  • Quantum Dots*
  • Tellurium
  • X-Rays

Substances

  • Cadmium Compounds
  • Gold
  • Tellurium
  • cadmium telluride