Postlingually Deafened Adult Cochlear Implant Users With Prolonged Recovery From Neural Adaptation at the Level of the Auditory Nerve Tend to Have Poorer Speech Perception Performance

Ear Hear. 2022 Nov-Dec;43(6):1761-1770. doi: 10.1097/AUD.0000000000001244. Epub 2022 Oct 18.

Abstract

Objective: This study investigated the effects of two temporal response properties of the auditory nerve (i.e., neural adaptation and recovery from neural adaptation) on speech perception performance in postlingually deafened adult cochlear implant (CI) users.

Design: Study participants included 18 postlingually deafened adults who were Cochlear Nucleus device users with a full electrode array insertion in the test ear(s). Neural adaptation and adaptation recovery of the auditory nerve (AN) were evaluated using electrophysiological measures of the electrically evoked compound action potential (eCAP). The amount of neural adaptation was quantified by the adaptation index within three time windows: 0 to 8.89 (window 1), 44.44 to 50.00 (window 2), and 94.44 to 100.00 ms (window 3). The speed of neural adaptation was estimated using a two-parameter power law function. To evaluate adaptation recovery of the AN, eCAPs to the last pulse of the 100-ms pulse train were recorded at masker-probe-intervals ranging from 1.054 to 256 ms in logarithmic steps. The amount of adaptation recovery was quantified by the adaptation recovery ratio. The time-constant of adaptation recovery was estimated using an exponential function with up to three components. Speech perception performance was evaluated by measuring consonant-nucleus-consonant (CNC) word scores presented in quiet and in speech-shaped noise at a signal-to-noise ratio (SNR) of +10 dB. One-tailed Pearson Product Moment correlation tests were used (1) to assess the associations among parameters of neural adaptation and adaptation recovery and (2) to evaluate the strength of association between these parameters and CNC word scores measured in quiet and in noise. The contributions of different parameters quantifying neural adaptation and adaptation recovery on speech perception scores were evaluated using multivariable linear regression analyses.

Results: The Pearson Product Moment correlation coefficient demonstrated a moderate, negative correlation between the speed of adaptation recovery and CNC word scores measured in quiet and in noise. The speed of adaptation recovery accounted for 14.1% of variability in CNC word scores measured in quiet and 16.7% of variability in CNC word scores measured in noise. The correlation strengths between CNC word scores and the adaptation index, the adaptation recovery ratio and the speed of neural adaptation ranged from negligible to weak.

Conclusions: The speed of adaptation recovery plays a more important role than other features of neural adaptation and adaptation recovery of the AN in speech perception in postlingually deafened adult CI users. Patients with prolonged adaptation recovery tend to show poorer speech perception performance.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Adult
  • Cochlear Implantation*
  • Cochlear Implants*
  • Cochlear Nerve
  • Humans
  • Noise
  • Speech Perception* / physiology