Simultaneous multislice diffusion-weighted imaging versus standard diffusion-weighted imaging in whole-body PET/MRI

Eur Radiol. 2023 Apr;33(4):2536-2547. doi: 10.1007/s00330-022-09275-4. Epub 2022 Dec 3.

Abstract

Objective: To compare standard (STD-DWI) single-shot echo-planar imaging DWI and simultaneous multislice (SMS) DWI during whole-body positron emission tomography (PET)/MRI regarding acquisition time, image quality, and lesion detection.

Methods: Eighty-three adults (47 females, 57%), median age of 64 years (IQR 52-71), were prospectively enrolled from August 2018 to March 2020. Inclusion criteria were (a) abdominal or pelvic tumors and (b) PET/MRI referral from a clinician. Patients were excluded if whole-body acquisition of STD-DWI and SMS-DWI sequences was not completed. The evaluated sequences were axial STD-DWI at b-values 50-400-800 s/mm2 and the apparent diffusion coefficient (ADC), and axial SMS-DWI at b-values 50-300-800 s/mm2 and ADC, acquired with a 3-T PET/MRI scanner. Three radiologists rated each sequence's quality on a five-point scale. Lesion detection was quantified using the anatomic MRI sequences and PET as the reference standard. Regression models were constructed to quantify the association between all imaging outcomes/scores and sequence type.

Results: The median whole-body STD-DWI acquisition time was 14.8 min (IQR 14.1-16.0) versus 7.0 min (IQR 6.7-7.2) for whole-body SMS-DWI, p < 0.001. SMS-DWI image quality scores were higher than STD-DWI in the abdomen (OR 5.31, 95% CI 2.76-10.22, p < 0.001), but lower in the cervicothoracic junction (OR 0.21, 95% CI 0.10-0.43, p < 0.001). There was no significant difference in the chest, mediastinum, pelvis, and rectum. STD-DWI detected 276/352 (78%) lesions while SMS-DWI located 296/352 (84%, OR 1.46, 95% CI 1.02-2.07, p = 0.038).

Conclusions: In cancer staging and restaging, SMS-DWI abbreviates acquisition while maintaining or improving the diagnostic yield in most anatomic regions.

Key points: • Simultaneous multislice diffusion-weighted imaging enables faster whole-body image acquisition. • Simultaneous multislice diffusion-weighted imaging maintains or improves image quality when compared to single-shot echo-planar diffusion-weighted imaging in most anatomical regions. • Simultaneous multislice diffusion-weighted imaging leads to superior lesion detection.

Keywords: DWI; Diffusion magnetic resonance imaging; Echo-planar imaging; Multiband; Parallel imaging.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Diffusion Magnetic Resonance Imaging* / methods
  • Echo-Planar Imaging / methods
  • Female
  • Humans
  • Magnetic Resonance Imaging
  • Male
  • Middle Aged
  • Positron-Emission Tomography* / methods
  • Reproducibility of Results
  • Whole Body Imaging* / methods