Background: Echocardiography-based transvalvular mean pressure gradient (ECHO-mPG) used to assess the forward valve function and structural valve deterioration could overestimate the true pressure gradient. This study evaluated the discrepancy between invasive and ECHO-mPG after transcatheter aortic valve implantation (TAVI) with respective valve type and size, its impact on a device success criterion, and predictors of a pressure discrepancy.
Methods: We analyzed 645 patients registered in a multicenter TAVI registry (balloon-expandable valve [BEV]: 500; self-expandable valve [SEV]: 145). The invasive transvalvular mPG was measured after valve implantation using two Pigtail catheters (CATH-mPG), while the ECHO-mPG was measured within 48 h after TAVI. Pressure recovery (PR) was calculated using the following formula: ECHO-mPG × effective orifice area (EOA)/ascending aortic area (AoA) × (1 - EOA/AoA).
Results: ECHO-mPG was weakly correlated with (r = 0.29, p < 0.0001), and consistently overestimated CATH-mPG in both BEV and SEV, and respective valve sizes. The magnitude of the discrepancy was larger for BEV than SEV (p < 0.001) and smaller valves (p < 0.001). After the correction of PR using the above formula, the pressure discrepancy remained for BEV (p < 0.001) but not SEV (p = 0.10). The proportion of patients with an ECHO-mPG > 20 mmHg decreased from 7.0% to 1.6% after correction (p < 0.0001). Among the baseline and procedural variables, post-procedural ejection fraction, BEV versus SEV, and smaller valves were associated with a larger discrepancy in mPG.
Conclusions: ECHO-mPG could be overestimated after TAVI, especially in patients with a smaller BEV. A higher ejection fraction, BEV, and smaller valves were predictors of a pressure discrepancy between CATH- and ECHO-mPG.
Keywords: Bernoulli equation; Pressure gradient; Pressure recovery; Transcatheter heart valve.