This study aimed to evaluate agreement of Wide scan measurements from swept-source optical coherence tomography(SS-OCT) Triton and spectral-domain OCT(SD-OCT) Maestro in normal/glaucoma eyes, and to assess the precision of measurements from Wide and Cube scans of both devices. Three Triton and three Maestro operator/device configurations were created by pairing three operators, with study eye and testing order randomized. Three scans were captured for Wide (12mm×9mm), Macular Cube (7mmx7mm-Triton; 6mmx6mm-Maestro), and Optic Disc Cube (6mmx6mm) scans for 25 normal eyes and 25 glaucoma eyes. Thickness of circumpapillary retinal nerve fiber layer(cpRNFL), ganglion cell layer+inner plexiform layer(GCL+), and ganglion cell complex(GCL++) was obtained from each scan. A two-way random effect analysis of variance model was used to estimate the repeatability and reproducibility; agreement was evaluated by Bland-Altman analysis and Deming regression. Precision limit estimates were low: <5µm for macular and <10µm for optic disc parameters. Precision for Wide and Cube scans of both devices were comparablein both groups. Excellent agreement between the two devices was found for Wide scans, with the mean difference<3µm across all measurements (cpRNFL<3µm, GCL+<2µm, GCL++<1µm), indicating interoperability. A single Wide scan covering the peripapillary and macular regions may be useful for glaucoma management.