Short-term safety and effectiveness of the mCLIP partial prosthesis

Eur Arch Otorhinolaryngol. 2024 May;281(5):2353-2363. doi: 10.1007/s00405-023-08359-1. Epub 2023 Dec 22.

Abstract

Purpose: This multicentric, retrospective study aimed to analyze the short-term safety and effectiveness of the mCLIP Partial Prosthesis.

Methods: Patients underwent tympanoplasty with implantation of a mCLIP Partial Prosthesis. Follow-up examination included ear microscopy and pure-tone audiometry to determine the post-operative pure tone average of the frequencies 0.5, 1, 2 and 3 kHz (PTA4). The post-operative PTA4 air bone gap (ABG) was used to evaluate the audiological outcome. A post-operative minimum and maximum follow-up period was not defined. Thus, the follow-up times of each study center were different, which resulted in different follow-up times for the audiological analysis and for adverse events (AE).

Results: 72 (66 adults, 6 children) patients were implanted with the mCLIP Partial Prosthesis. 68 (62 adults, 6 children) patients underwent audiological examination; all 72 patients were examined for adverse events. All patients (N = 68): 72.1% of the patients showed a PTA4 ABG of ≤ 20 dB. Individual post-operative bone conduction (BC) PTA4 thresholds were stable in 67 patients. The mean post-operative follow-up time was 78 ± 46 days. Children (N = 6): 5 out of 6 children showed a PTA4 ABG of ≤ 20 dB. None of the children reported a BC PTA4 deterioration of > 10 dB HL after the implantation. The mean post-operative follow-up time was 101 ± 45 days. Adverse events (all patients, N = 72): 15 (14 adults, 1 child) patients had AEs (27 AEs and 2 Follow-Ups). The mean post-operative follow-up time was 375 days.

Conclusion: Clinical data show satisfactory audiological parameters after implantation of the mCLIP Partial Prosthesis. The prosthesis is safe and effective for implantation in children and adults.

Trial registration number: NCT05565339, 09 September 2022, retrospectively registered.

Keywords: Hearing restoration; Ossiculoplasty; Passive middle ear implant; mCLIP partial prosthesis.

Publication types

  • Multicenter Study

MeSH terms

  • Adult
  • Audiometry, Pure-Tone
  • Bone Conduction
  • Child
  • Follow-Up Studies
  • Humans
  • Ossicular Prosthesis*
  • Prosthesis Implantation
  • Retrospective Studies
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT05565339