Data comparing MANTA device with Perclose device for large bore arterial access closure is limited. We performed meta-analysis to compare safety and efficacy of the two devices in large (⩾14 Fr sheath) arteriotomy closure post-TAVR. Relevant studies were identified via PubMed, Cochrane, and EMBASE databases until June, 2022. Data was analyzed using random effect model to calculate relative odds of VARC-2 defined access-site complications and short-term (in-hospital or 30-day) mortality. A total of 12 studies (2 RCT and 10 observational studies) comprising 2339 patients were included. The odds of major vascular complications (OR 0.99, 95% CI 0.51-1.92; p = 0.98); life threatening and major bleeding (OR 0.77, 95% CI 0.45-1.33; p = 0.35); minor vascular complications (OR 1.37, 95% CI 0.63-2.99; p = 0.43); minor bleeding (OR 0.94, 95% CI 0.57-1.56; p = 0.82); device failure (OR 0.74, 95% CI 0.49-1.11; p = 0.14); hematoma formation (OR 0.76, 95% CI 0.33-1.75; p = 0.52); dissection, stenosis, occlusion, or pseudoaneurysm (OR 1.08, 95% CI 0.71-1.62; p = 0.73) and short-term mortality (OR 1.01, 95% CI 0.55-1.84; p = 0.98) between both devices were similar. MANTA device has a similar efficacy and safety profile compared to Perclose device.
Keywords: Cardiology; MANTA device; Perclose device; closure device; interventional cardiology; large bore access; plug-based closure device; suture-based closure device; vascular medicine.