An IROA Workflow for correction and normalization of ion suppression in mass spectrometry-based metabolomic profiling data

Res Sq [Preprint]. 2024 Feb 1:rs.3.rs-3914827. doi: 10.21203/rs.3.rs-3914827/v1.

Abstract

Ion suppression is a major problem in mass spectrometry (MS)-based metabolomics; it can dramatically decrease measurement accuracy, precision, and signal-to-noise sensitivity. Here we report a new method, the IROA TruQuant Workflow, that uses a stable isotope-labeled internal standard (IROA-IS) plus novel companion algorithms to 1) measure and correct for ion suppression, and 2) perform Dual MSTUS normalization of MS metabolomic data. We have evaluated the method across ion chromatography (IC), hydrophilic interaction liquid chromatography (HILIC), and reverse phase liquid chromatography (RPLC)-MS systems in both positive and negative ionization modes, with clean and unclean ion sources, and across different biological matrices. Across the broad range of conditions tested, all detected metabolites exhibited ion suppression ranging from 1% to 90+% and coefficient of variations ranging from 1% to 20%, but the Workflow and companion algorithms were highly effective at nulling out that suppression and error. Overall, the Workflow corrects ion suppression across diverse analytical conditions and produces robust normalization of non-targeted metabolomic data.

Publication types

  • Preprint