Background: Atrial fibrillation (AF) recurrence after pulmonary vein isolation (PVI) is predominantly attributed to pulmonary vein reconnection (PVR). Predictors of AF recurrence have been widely studied; however, data are scarce on procedural parameters that predict chronic PVR. We aimed to study PVR rates and predictors of PVR. Methods: We retrospectively included 100 patients who underwent repeated ablation due to AF recurrence after initial PVI with the CARTO system. PVR was determined during the repeated procedure by electrophysiological evaluation, and initial procedural characteristics predicting PVR were studied, including adherence to the CLOSE protocol, use of high power, first-pass isolation (FPI), and baseline generator impedance (BGI). Results: Thirty-eight patients underwent initial CLOSE-guided PVI, and sixty-two underwent initial non-CLOSE PVI. A repeat procedure was performed 23 ± 16 months after the initial procedure. In total, PVR was found in 192 of 373 PVs (51.5%), and all PVs were isolated in 17/100 (17%) patients. Factors associated with all PVs being isolated were adherence to the CLOSE protocol, a higher power setting, the presence of bilateral FPI, and lower BGI (88% vs. 28%, p < 0.0001; 37.5 W vs. 30 W, p = 0.0276; 88.2% vs. 40.4%, p = 0.0007; and 127.6 Ω vs. 136.6 Ω, p = 0.0027, respectively). In initial procedures with adherence to the CLOSE protocol, the FPI rate was significantly higher (73.7% vs. 25%, p < 0.0001), while there were no significant differences in terms of procedure time and left atrial dwell time (81 vs. 85 min, p = 0.83; and 60 vs. 58 min, p = 0.08, respectively). BGI ≥ 130 Ω (AUC = 0.7403, sensitivity: 77.1%, specificity: 68.8%, p = 0.0032) was associated with a significantly higher probability of PVR (OR = 6.757; p < 0.0001). In multivariable analysis, independent predictors for PVR were non-adherence to the CLOSE protocol and BGI ≥ 130 Ω. Conclusions: Our findings indicate that adherence to the CLOSE protocol and baseline generator impedance < 130 Ω during AF ablation are independent predictors of PVI durability.
Keywords: CLOSE protocol; generator impedance; pulmonary vein isolation; pulmonary vein reconnection.