A slope method for the determination of electron energy for quality assurance

J Appl Clin Med Phys. 2024 Jul;25(7):e14369. doi: 10.1002/acm2.14369. Epub 2024 Apr 29.

Abstract

Background: Particle accelerators, manufactured for delivering patient radiation treatment, require numerous and frequent quality assurance measures. One of those is the periodic check for electron energy stability. The American Association of Physicists in Medicine has established requirements for this procedure. The current recommendation is to perform a ratio of two ionization points, one at Dmax and another at a point approximately to the 50% depth, compared to a baseline as a relative check.

Purpose: This ratio method is a sensitive measurement and sometimes produces results that are difficult to interpret or relate to acceptable tolerances. We sought to find a simple method that gives more stable results, which can be interpreted and related to energy changes.

Method: We propose a method that takes two measurements on the descending portion of the shifted percent depth ionization (PDI) curves to calculate the slope, tangent to the I50 point, the point at which the ionization falls to 50% of its maximum value. We then used the slope measurement, compared to an established baseline, to relate energy.

Results: After collecting data over a 3-year period, we saw that standard deviations for the slope method have much less variability than the traditional ratio method. We were also able to correlate the slope results to ionization scans performed in water and found they were in better agreement than the traditional ratio method.

Conclusion: The slope method does not require precise positioning since the slope remains relatively constant over the descending portion of the curve. Our data show that this results in an easier interpretative test of electron energy stability and delivers reliable feedback for quality assurance.

Keywords: TG‐141; TG‐198; electron beam energy check; ionization; monthly quality assurance.

MeSH terms

  • Electrons*
  • Humans
  • Particle Accelerators* / instrumentation
  • Particle Accelerators* / standards
  • Quality Assurance, Health Care* / standards
  • Quality Control
  • Radiometry / methods
  • Radiometry / standards
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods