Hypothesis: Resonance frequency analysis (RFA) is a reliable, noninvasive method to assess the stability of bone-anchored hearing implants (BAHIs), although surgical-, implant-, and host-related factors can affect its outcome.
Background: BAHI plays an important role in restoring hearing function. However, implant- and host-related factors contribute to premature implant extrusion. To mitigate this, noninvasive methods to assess implant stability, along with a better understanding of factors contributing to BAHI failure, are needed.
Methods: We evaluated the utility of RFA to quantify implant stability in sawbone (bone mimicking material), 29 human cadaveric samples, and a prospective cohort of 29 pediatric and 27 adult participants, and identified factors associated with implant stability. To validate the use of RFA in BAHI, we compared RFA-derived implant stability quotient (ISQ) estimates to peak loads obtained from mechanical push-out testing.
Results: ISQ and peak loads were significantly correlated (Spearman rho = 0.48, p = 0.0088), and ISQ reliably predicted peak load up to 1 kN. We then showed that in cadaveric samples, abutment length, internal table bone volume, and donor age were significantly associated with implant stability. We validated findings in our prospective patient cohort and showed that minimally invasive Ponto surgery (MIPS; versus linear incision), longer implantation durations (>16 wk), older age (>25 yr), and shorter abutment lengths (≤10 mm) were associated with better implant stability. Finally, we characterized the short-term reproducibility of ISQ measurements in sawbone and patient implants.
Conclusions: Together, our findings support the use of ISQ as a measure of implant stability and emphasize important considerations that impact implant stability, including surgical method, implant duration, age, and abutment lengths.