Background: Plasma levels of D-dimer are elevated in patients with thromboembolisms. Here we investigated the existence of interfering antibodies as a potential cause for elevated D-dimer levels.
Case presentation: A 42-year-old white Caucasian woman with a prior history of pulmonary embolism during her first pregnancy (treated with heparin therapy for 6 weeks postnatally) and hypothyroidism had a persistent elevated D-dimer without any clinical or ultrasound-based signs of thromboembolic conditions during her second pregnancy. We obtained informed consent and plasma was obtained from the patient. D-dimer levels were measured using two different assays. We also tested for the presence of rheumatoid factor, performed dilution series, and finally used an antibody depletion strategy. The two D-dimer assays performed similarly. Using our antibody depletion technique, we observed that ~ 1/3 of the increased plasma levels of D-dimer may be attributed to interfering antibodies.
Conclusions: Our results identify interfering antibodies as a potential contributor to an increased D-dimer in this patient. Our case highlights the potential of heterophilic interference for increased D-dimer and provides a procedure to determine this analytically.
Keywords: d-dimer; Case report; Heterophilic antibodies; Immunoassay; Interfering antibodies.
© 2024. The Author(s).