Improvement of resolution in measurement of electrocardiogram RR intervals by interpolation

Med Eng Phys. 1997 Jun;19(4):375-9. doi: 10.1016/s1350-4533(96)00067-7.

Abstract

The measurement of successive RR intervals, obtained from the electrocardiogram (ECG), is the basis of any subsequent method for the assessment of variations in heart rate, for example using intervalograms, histograms, trend curves, spectral analysis, etc. The accuracy of measurement directly depends on the sampling rate used for the acquisition of the ECG signal. Depending on specific applications, rates down to 128 Hz are not uncommon. This considerably limits the resolution if RR interval data are to be derived. Interpolation is often employed for solving problems of this type. Linear, cubic and spline interpolations, obtained directly from the MATLAB software product, were applied and compared for the purpose of RR interval measurement. The cubic method was found to combine the improvement of resolution (+/-1 ms) at sampling rates successfully down to 100 Hz, with a rapid operating speed. Comparison examples are given with ECG signals acquired with 1000 Hz, 16-bit sampling.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't
  • Technical Report

MeSH terms

  • Adult
  • Aged
  • Databases, Factual
  • Electrocardiography / instrumentation
  • Electrocardiography / methods*
  • Electrocardiography / statistics & numerical data
  • Electrodes
  • Female
  • Humans
  • Male
  • Middle Aged
  • Signal Processing, Computer-Assisted