The benefits shown by the recent introduction of PCR for the in vitro diagnosis of hepatitis C virus (HCV) infection has prompted the development of standardized, ready-to-use assays that can be implemented in routine clinical laboratories. We have evaluated the clinical performance of COBAS AMPLICOR HCV (COBAS), the first instrument system that allows the automation of HCV RNA amplification and detection, to determine its performance in the routine laboratory setting. More than 2,000 specimens collected at five centers were analyzed in parallel by the COBAS and the manual AMPLICOR HCV (AMPLICOR) tests, and the results were compared with the results for biochemical and serological markers of HCV. In this study the two PCR systems showed the same accuracy, with a concordance rate of 99.8%. As expected, the correlation between serology and PCR was not absolute because the presence of anti-HCV antibodies may be associated with a latent or past infection. On the other hand, if the presence of confirmed anti-HCV antibodies and elevated alanine aminotransferase levels are taken as the "gold standard," indicating an active, ongoing infection, the COBAS and AMPLICOR tests show high and comparable sensitivities (100%) and specificities (98%), with positive and negative predictive values of 100 and 97%, respectively. During the study no false-positive reactions were detected. The use of an internal control allowed the identification of inhibitory substances that prevented amplification for 0.3 and 0.4% of samples tested by the COBAS and AMPLICOR tests, respectively. Compared to the manual system, the COBAS system allowed a significant reduction of hands-on time and could improve the overall laboratory work flow. In conclusion, these results support the use of the COBAS and AMPLICOR tests for the molecular diagnosis of active HCV infections.