Bilirubin is a by-product of heme catabolism [
1], where less than 1.2 mg/dL is always present in the blood of healthy individuals [
2]. It exists as conjugated and unconjugated bilirubin (UCB) in the bloodstream. Unconjugated bilirubin (UCB) is toxic and water-insoluble; hence, it should be excreted. The excretion starts with UCB binding with albumin and is transported to the liver. At this point, UCB is enzymatically converted into conjugated bilirubin (CB). The latter is less toxic and water-soluble, allowing for renal or intestinal excretion [
3]. However, in the case of pathophysiological events, bilirubin levels are elevated, and the liver fails to live up to its conjugation ability, leading to irreversible neurological damage or death [
4]. High bilirubin levels have been correlated with hepatic and hemolytic disorders [
5,
6].
Several studies on BR levels estimation have been reported [
7,
8], but these techniques suffer from the instrument complexity and cost. This mini-paper investigated the dual-wavelength (DWL) technique for BR estimation using a system-on-chip (SoC).
2. Materials and Methods
Materials: Bilirubin (PN: 14370), Dimethyl Sulfoxide (DMSO) and sodium citrate (NaCHO) commercial standards were supplied from Sigma-Aldrich, Australia.
Blood sample preparation: Porcine blood was procured from the abattoir immediately after sacrifice and was preferred as a prominent replacement for human blood due to its biochemical resemblance [
9] and low cost. Solutions were prepared by mixing BR at varied concentrations and blood with anticoagulant (4%
w/
v)-to-blood ratio of 1:9 (
v/
v) [
10]. BR-blood samples were stored at ~5 °C in the dark to prevent photodegradation. For extended preparation steps, refer to the study performed by Ndabakuranye et al. [
11].
Optical measurement: A SoC platform was used for BR measurement. The setup consisted of a MAX86916 optical module, a MAX32630 host and an optical stage designed to operate in transflection mode. Transflection mode implies that a rear optical reflector (E39-R42, Omron) was used on top of the sample. Therefore, the spectral response contained both light intensity backscattered from the sample and light transmitted through the sample then reflected by the reflector to be retransmitted back through the sample, thereby doubling the path length (
Figure 1). Major features of the system-on-chip are summarized in
Table 1.
Feature selection: Several features were optimally selected using theoretical investigations and simulations to ensure the accuracy, reliability, and safety of bilirubin measurements. These features are summarized in
Table 2.
3. Results and Discussions
The analytical (470 nm) and reference (525 nm) wavelengths were obtained by analyzing the distinct optical signatures of blood and BR. It does not significantly absorb green light but strongly absorbs blue light (ε
525 = 214, ε
470 = 46188 [cm
−1 M
−1]) [
12]. However, although the LEDs on the SoC emit the maximum power at the wavelengths of 455 nm and 530 nm, the LEDs’ emission bands are sufficiently wide to also cover 470 nm and 525 nm, since the LEDs are not ideal monochromatic sources.
The DWL measurement was performed using a revised 2-dimensional Beer’s law as shown in Equation (1) which suggests that BR concentration can be correlated with its absorbance ratio at 470 and 525 nm (
t: path length [cm],
ε: extinction coefficient,
C: concentration).
To investigate
BR’s sensitivity in blood,
R-parameters were deduced by collecting the MAX86916 ADC counts data and calculated using Equation (2).
Figure 1 shows the SoC platform used to measure the ADC data used to compute
R-parameters (
). The plot of R-parameters as a function of
BR concentrations is shown in
Figure 2, and results showed a strong linear relationship with the coefficient of determination greater than 0.991.
Although this technique provides a robust and simple way of measuring BR, it may be susceptible to errors due to LEDs’ spectral and spatial distribution inadequacies, residual bilirubin, other hemoglobin forms (COHb or MetHb), SaO2 and Hb levels variability.
4. Conclusions
The feasibility of BR monitoring by the DWL method was investigated, and BR’s sensitivity in blood was explored at pathophysiological ranges (1.2–50 mg/dL) using an Soc. The SoC includes a miniature MAX86916 optical module with integrated signal conditioning and processing capabilities. Results showed a strong correlation between R-parameters and BR concentration. Our findings lay the groundwork for point-of-care testing for BR levels, primarily for hyperbilirubinemia infants and cirrhotic adults out of clinical settings.
Author Contributions
Methodology, J.P.N.; Formal Analysis, J.P.N.; Investigation, J.P.N.; Data Curation, J.P.N.; Original Draft Preparation and Writing, J.P.N.; Conceptualisation, J.P.N., A.A. and S.P.; Writing—review and editing, J.P.N., A.A. and S.P.; Supervision, S.P. and A.A.; Project Administration, S.P. and A.A.; Funding Acquisition, S.P. and A.A. All authors have read and agreed to the published version of the manuscript.
Funding
This research was funded by the Australian Research Council, grant number LP160101052.
Institutional Review Board Statement
Experiments were carried out on samples from the abattoir, and thus, neither ethics application nor written consent was required.
Acknowledgments
Jean Pierre gratefully acknowledges the Melbourne Research Scholarship (the University of Melbourne), The School of Physics (The University of Melbourne) and the STEM college (RMIT University).
Conflicts of Interest
Authors declare that they have no conflicting financial or non-financial interests.
References
- Ahmad, H.B.; Ahmad, S.; Shad, M.A.; Hussain, M. Kinetic measurements for photodecomposition of bilirubin. Asian J. Chem. 2013, 25, 7945–7948. [Google Scholar] [CrossRef]
- Dufour, D.R.; Lott, J.A.; Nolte, F.S.; Gretch, D.R.; Koff, R.S.; Seeff, L.B. Diagnosis and monitoring of hepatic injury. I. Performance characteristics of laboratory tests. Clin. Chem. 2000, 46, 2027–2049. [Google Scholar] [CrossRef] [PubMed]
- Berk, P.D. Bilirubin metabolism and the hereditary hyperbilirubinemias. Semin. Liver Dis. 1994, 14, 321–322. [Google Scholar] [CrossRef] [PubMed]
- AAP. Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics 2004, 114, 297–316. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Lopez-Velazquez, J.A.; Chavez-Tapia, N.C.; Ponciano-Rodriguez, G.; Sanchez-Valle, V.; Caldwell, S.H.; Uribe, M.; Mendez-Sanchez, N. Bilirubin alone as a biomarker for short-term mortality in acute-on-chronic liver failure: An important prognostic indicator. Ann. Hepatol. 2014, 13, 98–104. [Google Scholar] [CrossRef]
- D’Amico, G.; Garcia Tsao, G.; Pagliaro, L. Natural history and prognostic indicators of survival in cirrhosis: A systematic review of 118 studies. J. Hepatol. 2006, 44, 217–231. [Google Scholar] [CrossRef] [PubMed]
- Hulzebos, C.V.; Vitek, L.; Zabetta, C.D.C.; Dvořák, A.; Schenk, P.; van der Hagen, E.A.; Cobbaert, C.; Tiribelli, C. Diagnostic methods for neonatal hyperbilirubinemia: Benefits, limitations, requirements, and novel developments. Pediatr. Res. 2021, 90, 277–283. [Google Scholar] [CrossRef] [PubMed]
- Rawal, R.; Kharangarh, P.R.; Dawra, S.; Tomar, M.; Gupta, V.; Pundir, C. A comprehensive review of bilirubin determination methods with special emphasis on biosensors. Process Biochem. 2020, 89, 165–174. [Google Scholar] [CrossRef]
- Pond, W.G.; Mersmann, H.J. Biology of the Domestic Pig; Comstock Pub. Associates, Cornell University Press: Ithaca, NY, USA, 2001. [Google Scholar]
- Sorapukdee, S.; Narunatsopanon, S. Comparative study on compositions and functional properties of porcine, chicken and duck blood. Korean J. Food Sci. Anim. Resour. 2017, 37, 228–241. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Ndabakuranye, J.P.; Rajapaksa, A.E.; Burchall, G.; Li, S.; Prawer, S.; Ahnood, A. A novel optical assay system for bilirubin concentration measurement in whole blood. IEEE Trans. Biomed. Eng. 2021, in press. [CrossRef]
- Yao, J.; Wang, L.V. Photoacoustic microscopy. Laser Photonics Rev. 2013, 7, 758–778. [Google Scholar] [CrossRef]
| Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).